Electrode interface design
Developed the flex PCB in KiCad around the array, flex lead, connector, and channel mapping requirements.
NEURAL INTERFACES × HARDWARE
Build the recording interface that the experiment needs.
THE IDEA
In the Neuromechatronics Lab at Carnegie Mellon University, advised by Douglas Weber, I designed a 32-channel high-density EMG flex PCB with polyimide and FR-4 stiffeners. The design combines an 8 × 4 electrode array at 8 mm pitch, a 300 mm flex lead, and an Omnetics connector.
MY CONTRIBUTION
Developed the flex PCB in KiCad around the array, flex lead, connector, and channel mapping requirements.
Used Python to automate component and layout generation, making repeated geometry and channel assignments easier to manage.
Verified channel connectivity in the design files, worked through the fabricator’s DFM/DFA review, prepared the assembly drawing, and approved the production Gerbers.
ENGINEERING CHALLENGE
For an electrode array, spatial arrangement and electrical mapping must agree. My work combines layout automation with connectivity checks so the design can be reviewed as a recording interface, with a clear mapping from electrode locations to connector channels.
RESULTS & SCOPE
32 channels in an 8 × 4 array at 8 mm pitch, with 4.22 mm electrode pads.
300 mm flex lead at 0.10 mm trace and space, ending in an Omnetics connector.
Zero DRC and ERC violations in KiCad; fabricator DFM/DFA review completed; 16 units ordered and boards received.
The boards have been received but not yet tested. Fabrication yield, recording quality, and in-use performance are not claimed here. The illustration on this page is a conceptual array diagram, not a manufacturing drawing.